Intelligent lighting incubator instructions

I. Light incubator features:
The light incubator has an over-temperature and sensor abnormal protection function to ensure the safety of the instrument and the sample; the plant full-spectrum plant growth lamp is selected to facilitate plant growth and improve disease resistance. With power-down memory, power-down time automatic compensation function; constant temperature control system, fast response, high temperature control accuracy.

Second, the light incubator uses:

The light incubator microcomputer fully automatic control, touch switch, easy to operate; programmable multi-segment control mode, temperature, humidity, illuminance and time can be set separately during day and night. Air duct ventilation, studio wind speed is gentle, temperature is uniform; hollow reflective tempered film glass, good thermal insulation performance, beautiful appearance. Fully enclosed opaque lampshade, optional studio power supply, disinfection device, etc.

Third, the light incubator technical parameters:

1. Volume: 150L 250L

2. Temperature control range: 0~50°C

3. Temperature fluctuation: ±0.5 °C

4. Temperature uniformity: ± 1 ° C

5. Illuminance: 0-3500-5000LX

Light can be selected, for example: seed germination can be selected 3500LX, seedling growth can be selected 7500LX or above. All incubators in our company can increase the light intensity according to user requirements.

Fourth, the light incubator instructions:

1. Turn on the power, close the power switch, power on the whole machine, and the power indicator in the switch is on.

2. Please refer to the manual of intelligent digital temperature controller for temperature control setting.

3. If you need lighting, turn on the light switch.

4. This incubator has a breakpoint protection function and a delay function of about one and a half minutes. It takes about one and a half minutes to start again after the compressor stops.
5. Maintenance and cultivation of light incubator:

1. The incubator housing should be reliably grounded.

2. The incubator should be placed in a cool, dry, well-ventilated place away from heat and sunlight. Place it smoothly to prevent noise from occurring.

3. In order to ensure the effective heat dissipation of the condenser, the distance between the condenser and the wall should be greater than 100mm. The side of the box should have a gap of 50mm, and the top of the box should have at least 300mm space.

4. When the incubator is transported, repaired and maintained, it should avoid collision, shaking and shaking; the maximum inclination is less than 45 degrees.

5. The instrument suddenly does not work. Please check whether the fuse tube (after the box) is burnt out and check the power supply.

6. When the incubator is working in refrigeration, it is not advisable to make the difference between the temperature inside the box and the ambient temperature greater than 25 degrees.
Common faults of the light incubator 1. The compressor is not cooled, the temperature inside the incubator is not uniform, or the circulating air outlet is frozen.

2. In the balanced mode, the compressor is always in working condition, the cooling effect is better and the compressor can be effectively maintained.

3. Uneven temperature:

4, the start is easy to shorten its service life.

5. If the temperature is reduced to a certain value if it is a single-mode operation, it will work when the temperature is lowered. When the door is opened or closed, the temperature will fluctuate greatly. Causing the compressor to be ordinary

6. It may be caused by excessively dense discharge items or the circulating fan inside the box does not work.

7. If you use long light, the light bulb will easily heat up for a long time, which will cause uneven temperature.

The outlet is frozen:

Each refrigerated instrument has a defrost cycle during which the compressor does not cool. The ice on the condenser is melted by the heating element. The melted water is likely to collect at the air outlet and eventually cause ice to form.